Product Description
Fabricated under the supervision of our qualified engineers using premium grade components, this Fully Automated Bio-Chemistry Analyser is extensively used in hospitals, research centers and laboratories. Offered analyser is widely appreciated among our patrons and used for conducting 440 tests in an hour. Stringently checked on various parameters, this Fully Automated Bio-Chemistry Analyser can be availed from us in different models and other technical specifications as per the requirements of clients.
Features:
Modular design
Robust construction
Easy to operate
Fast results
Specifications:
NOVA 2200
NOVA 2200 Fully Automatic Bio-Chemistry Random Access Analyzers are high throughput and High Speed Auto Analyzers having throghput of 220 Tests/Hr and 440 Tests/Hour respectively. Our all products having in product quality accreditation like ISO-9001,ISO-14001,ISO- 13485,CE,WHO-GMP & FDA
Specification
- Throughput : 200 test / hour, up to 280 test/ hour with ISE (optional)
- Test Mode : End point, kinectic, fixed time, Immuno assays.
- Test Range : Clinical Biochemistry, Turbidmetry, ISE, Portain, Drug abuse
- Test Setting : Programmable for single, multi combined, batch items tests
- Parameter settings : Completed open for editing test parameters and calculate parameters
Sample System
- Sample tray : 44 sample positions can be extended to 88 sample positioning for primary tubes, test tibes and various sample cups.
- Whole tray can be rempved and replaced.
- Any position can be aligned as the Emergnecy test position.
- Sample volume : 2 ~30 ul, step by 0.1 ul
- Retest : Dilutes and re test auotmatically when sample concentration beyond limitation
- Sample probe : Liquid level detection and collision protection
- Probe cleaning : Automatic washing for both inside and outside,carry over <0.1 %
- Syringe : high performance syringe, endurable and manitenance free
- Optional Item : Bar code Reader
Reagent Systems
Reagent tray : 44 positions for single & double reagents positions can be aligned freely Refrigreration : Refrigerated reagent comparment, 24 hours non - stop cooling wither peltier element, maintaining temperature 4-15 degree CReagent botlle volume : 35 ml & 20 mlReagent volume : 150 ul - 300 ul, step by 1 ulReagent Needle : Automatic liquid level detection function & collision protectionProbe Cleaning : Automatic washing for bith inside and outside carry over < 0.1 %
Optical systems
- Light source : Halogen lamp 6 V/ 10 W
- life span > 3000 hours
- Wavelength : 9
- Interface filters : 340 nm, 405 nm, 450 nm, 505 nm, 546nm, 578nm, 620nm, 670nm
- Wavelenght accuracy : + - 2 nm
- Absorbance range : 0.000 - 3.000 Abs
- Resolution : 0.001 Abs
Calibration & Control
- Calibration mode : Linear (one point and multi point) Logit - log 5 P, Exponnential, Polynomial, parabola
- Control rules : Westgard multi rule, cumulative sum check, twin plot, 3 level s quality control
- Control mode : Real tome, intra day & inter day QC
Operating Systems
- Computer control : Windows 2000 systems or above
- Data management : Large capacity for results, reaction curve be displayed, saved and printed
- Interface : RS 323 interface
- Working Conditions Power supply : AC 220 V + - 22V, 50 Hz + - 1Hz
- Power consumption : < 500 VA
- Environmental temperature : 15 degree C ~ 30 degree C
- Humidity : < 85 %
- Water consumption : 5L/ hour
- Dimension : 760mm x 560 mm x 456 mm {LXWXH}
- Weight : 90 Kg